An intrauterine cavity morcellator: A novel approach to high volume uterus morcellation. Ex-vivo study by Meir Pomeranz

CONCLUSION: In this novel experiment, we verified the feasibility and safety of the Heracure device for vaginal intra-uterine morcellation for uterine size reduction. This technique could enable rapid and easy removal of the uterus through the vaginal orifice.

PLoS One. 2023 Mar 17;18(3):e0282149. doi: 10.1371/journal.pone.0282149. eCollection 2023.

ABSTRACT

OBJECTIVE: Uterine size is one of the essential factors determining the feasibility of a minimally invasive gynecologic surgery approach. A traditional electromechanical morcellator is a well-known tool but not without flaws. We aim to assess feasibility and safety of a novel intrauterine power morcellation device for uterine size reduction to overcome these limitations during hysterectomy.

METHODS: This single-arm, observational study was conducted in a single tertiary care medical center from April 2022 to July 2022. Feasibility and safety of a novel intrauterine morcellation device for uterine size reduction was tested in ten post-hysterectomy uteri (Ex-vivo).

MEASUREMENTS AND MAIN RESULTS: Ten uteri were examined in this trial. No major complications occurred during the procedure. All ten (10) uteri were successfully reduced in size (size reduction range was between 9% to 54%). The average resection time using the Heracure Device was 4.3 minutes (range: 1min- 10min). Mean uterus weight reduction was 21%, with a mean circumference reduction of 25%. No leakage was observed from the outer surface of the uterus/serosa after the saline injection post-procedure examination.

CONCLUSION: In this novel experiment, we verified the feasibility and safety of the Heracure device for vaginal intra-uterine morcellation for uterine size reduction. This technique could enable rapid and easy removal of the uterus through the vaginal orifice.

CLINICAL TRIAL REGISTRATION: Name of the registry: ClinicalTrials.gov; Number Identifier: NCT05332132.

PMID:36930660 | DOI:10.1371/journal.pone.0282149

Read More